Overview
Drop more than the reject threshold and the part leaks. The method is simple. Getting repeatable results out of it is not, and that comes down to knowing what actually moves the measurement.
Test Theory
The pressure decay rate is related to the leak rate by the Ideal Gas Law. For a fixed test volume at constant temperature:
- Leak rate (scc/min) = (ΔP × V) / (Patm × t) where ΔP is pressure change, V is volume, t is time
- Smaller test volumes provide higher sensitivity (less air to lose)
- Higher test pressures increase sensitivity proportionally
- Longer test times allow smaller leaks to be detected
- Typical sensitivity: 10⁻² to 10⁻⁴ scc/min depending on conditions
For air at standard conditions: Q (scc/min) = (ΔP × V) / (14.696 × t) where ΔP is in PSI, V is in cc, and t is in minutes. A 0.001 PSI drop in 100cc over 10 seconds = 0.041 scc/min leak rate.Test Cycle Phases
A pressure decay test runs in four phases: Fill (pressurize the part), Stabilize (let pressure and temperature settle), Test (measure the change), and Exhaust (vent it). Stabilize is the one that trips people up. Cut it short and thermal effects either hide a real leak or invent one that isn't there, so the result is wrong either way.
Factors Affecting Accuracy
Temperature is the biggest source of error. A 0.1°F swing in a 100cc volume at 30 PSI shows up as a pressure change the size of a real leak. Other culprits: adiabatic heating during fill (compressed air heats up), moisture evaporating off the surface, and flexible parts flexing under pressure.
Typical applications
- Automotive component testing
- Medical device seal verification
- Consumer product waterproofing
- HVAC component testing